GDOT-80A Insulation Oil Tester manual-updated1105
●Ndapota verenga bhuku rekushanda zvakanaka usati washanda.
●Ndokumbira utarise kana tester yakanyatsobatana nenyika isati yaedzwa.
●Zvinorambidzwa kufambisa kana kusimudza chivharo chekuedza mukati mekuedza kudzivirira kukuvadzwa nemhepo yakakwira.Simba rinofanirwa kudzima risati ratsiva sampling oil.
●Bata nekungwarira paunenge uchibvisa kana kuvhara high voltage test cover!
●Kana iyo tester ikashanda zvisingaite mushure mekunge oiri yekudzivirira yaparara, ndapota dzima tester kwemasekonzi gumi, wobva watanga zvakare.
●Mushure mekunge bepa rekudhinda rapera, ndapota tarisa kuprinta tsananguro chikamu (kana manual appendix) kutsiva bepa rekudhinda kudzivirira kukuvara kweprinta musoro.
●Chengetedza tester iri kure nekunyorova, huruva uye zvimwe zvinhu zvinoparadza, uye chengetedza iri kure kubva kune yakakwira tembiricha masosi.
●Bata nekungwarira mukufambisa.Usaise kumusoro pasi.
●Bhuku racho rinogona kudzokorora zvakakodzera pasina imwe ziviso mberi.Kana pane mubvunzo, tapota taura nesu.
Nguva yewaranti yeiyi nhevedzano igore rimwe kubva pazuva rekutumira.Ndokumbira utarise kune invoice yako kana magwaro ekutumira kuti uone akakodzera mazuva ewaranti.HVHIPOT corporation waranti kumutengi wekutanga kuti chigadzirwa ichi hachizove chisina hurema muzvinhu uye magadzirirwo pasi pekushandiswa kwakajairwa.Munguva yese yewaranti, ipai kuti kukanganisa kwakadaro hakuna kutsanangurwa neHVHIPOT kuve kwakakonzerwa nekushungurudzwa, kushandiswa zvisizvo, kuchinjwa, kuiswa zvisina kunaka, kuregeredza kana kushata kwemamiriro ekunze, HVHIPOT inongogumira chete kugadzirisa kana kutsiva chiridzwa ichi panguva yewaranti.
GDOT-80C chiridzwa | 1 pc |
Kapu yemafuta (250ml) | 1 pcs |
Tambo yemagetsi | 1 pc |
Spare fuse | 2 pcs |
Kubvongodza tsvimbo | 2 pcs |
Standard gauge (25mm) | 1 pc |
Prinda bepa | 2 rolls |
Tweezer | 1 pc |
Gaidhi remushandisi | 1 pc |
Factory test report | 1 pc |
HV Hipot Electric Co., Ltd. yakanyatso uye nekuchenjerera kuverenga bhuku rekushandisa, asi isu hatigone kuvimbisa kuti hapana zvikanganiso uye zvakasiiwa zvizere mubhuku rekushandisa.
HV Hipot Electric Co., Ltd. yakazvipira kuenderera mberi nekuvandudza mumabasa echigadzirwa, uye kuvandudza kunaka kwesevhisi, saka kambani inoramba iine kodzero yekuchinja chero zvigadzirwa uye mapurogiramu esoftware anotsanangurwa mubhuku rino pamwe nezviri mubhuku rino pasina kutanga. chiziviso.
Iko kuvharirwa kwemukati kweakawanda emidziyo yemagetsi mumasisitimu emagetsi, masisitimu enjanji, hombe-yakakura petrochemical zvidyarwa uye mabhizinesi, anonyanya kugamuchirwa nemafuta-akazadzwa nemhando yekuisa, saka, iyo insulating mafuta dielectric simba bvunzo yakajairika uye inofanirwa.Kuti tisangane nezvinodiwa pamusika, takagadzira uye kugadzira akateedzana ekudzivirira mafuta dielectric simba testers zvinoenderana nenyika standard GB/T507-2002, indasitiri yakajairwa DL429.9-91 uye yazvino Electric Power Indasitiri Standard DL/T846.7 -2004 isu pachedu.Ichi chiridzwa, nekushandisa single-chip microcomputer seyo musimboti, chakaitwa otomatiki, kuyerwa kwakaringana, kuvandudza zvakanyanya kushanda zvakanaka uye kuderedza kuwanda kwevashandi.Uyezve, idiki muhukuru uye yakanakira kutakura.
●Iine microprocessor, zadzisa otomatiki bvunzo yekumira yevoltage yekutenderera kweoiri ine huwandu hwe0 ~ 80KV (kusanganisira kuwedzera, kuchengetedza, kusanganisa, kumira, kuverenga, kudhinda uye mamwe maoperation).
●Yakakura LCD skrini inoratidza.
●Kushanda kuri nyore.Muchina uyu unozopedzisa otomatiki bvunzo dzekumira pane imwe kapu yemafuta emuenzaniso mushure mekugadzikwa kwakapusa nemushandisi.Kuparara kwemagetsi kukosha kwe1 ~ 6 nguva uye nguva dzekutenderera dzinozochengetwa otomatiki.Mushure mebvunzo, iyo inodhinda printer inodhinda yega yega yekuparara voltage kukosha uye avhareji kukosha.
●Simba-pasi kuchengetedza.Inogonachengetedza zana remhinduro uye ratidza tembiricha iripo uye hunyoro.
●Gamuchira single-chip microcomputer kuti ukwidze voltage nekumhanya nguva dzose.Iyo voltage frequency ndeyechokwadi pa50HZ, iva nechokwadi chekuti maitiro ese ari nyore kudzora.
●Ne-over-voltage, over-current and limit protection kuti ive nechokwadi chekuchengetedzwa kwevashandi.
●Nebasa rekuratidza tembiricha yakayerwa uye system wachi.
●Taurirana nekombuta neyakajairwa RS232 interface.
Magetsi | AC220V±10%, 50Hz |
Output voltage | 0~80kV(Selectable) |
Capacity | 1.5kVA |
Simba | 200W |
Speed yevoltage inowedzera | 2.0~3.5kV/s (Inochinjika) |
Voltagekuyerakururama | ±3% |
Waveform kukanganisa | ≤3% |
Kuwedzera nguva | 5min (inochinjika) |
Kumira nguva | 15min (inochinjika) |
Kuwedzera nguva | 1~6 (Inosarudzwa) |
Kushandazvakatipoteredza | Tkudziya: 0 ℃-45°C |
Humidity:Max.hunyoro hwehukama75% | |
Dimension | 465x385x425mm |
① Thermal printer--kudhinda mhinduro dzebvunzo;
② LCD - kuratidza menyu, kukurumidza uye bvunzo mhinduro;
③ Makiyi ekushandisa:
Dzvanya "◄" kiyi yekuwedzera kukosha kwekuseta;
Dzvanya "►" kiyi yekudzikisa kukosha kwekuseta;
Sarudza--yekusarudza mabasa (chinhu chakasarudzwa chiri pachiratidziro chekuchengetedza);
Simbisa--yekuita mabasa;
Kumashure--yekubuda mushandisirwo yekushandisa;
④ Simba switch uye chiratidzo
1. Kugadzirira kusati kwaedzwa
1.1 Batanidza nzvimbo yepasi (kurudyi rwemidziyo) kune waya yepasi zvakasimba usati washandisa midziyo.
1.2 Bvisa sampu yemafuta zvinoenderana neyakajairwa chiyero.Gadzirisa chinhambwe che electrode mukati mekapu yemafuta seyakajairwa geji.Geza mukombe maererano nezvinodiwa.Dururira sampu yemafuta mukapu uye uvhare chivharo.
1.3 Kuchinja kwemagetsi eAC220V mushure mekunge zvinhu zviri pamusoro zvasimbiswa, zvakagadzirira bvunzo.
2. Kuedza
2.1 Dzvanya switch yemagetsi wobva wapinda mune inotevera interface:
2.2 System Parameter kuseta
Dzvanya "Confirm" kiyi uye pinda mune inotevera interface:
Kusimudzira kuseta: vashandisi vanogona kusarudza zvinoenderana nezvinodiwa chaizvo.
Dzvanya "Back" kiyi yekubuda iyi interface mushure mekugadzirisa kwaitwa.
2.3 Kuedza
Dzvanya kiyi ye "Sarudza" kuti usarudze "Start Test" menyu uye tinya "Confirm" kiyi yekupinda mune inotevera interface:
Kuti uenderere mberi nebvunzo inotevera nekukurumidza kana bvunzo yekutanga yapera kusvika iyo seti yekuwedzera frequency yapera.Pakupedzisira, chigumisiro chinoratidzwa uye chakadhindwa sezvinotevera:
2.4 Kuona uye kudhinda data:
Dzvanya kiyi ye "Sarudza" kuti usarudze "Data Kuona uye Kudhinda" menyu uye tinya "Confirm" kiyi kuti upinde mune inotevera interface:
Sarudza "Peji Kumusoro" kana "Peji Pasi" uye sarudza zvinyorwa zvichadhindwa uye sarudza "Printa".
●Kusarudzwa kwemafuta emuoiri uye kuiswa kwechinhambwe che electrode kuchasangana neakakodzera nyika nemaindasitiri zviyero.
●Vashandisi kana vamwe vashandi vanorambidzwa zvachose kubata goko mushure mekunge simba rabatidzwa kudzivirira tsaona.
●Simba richagurwa nekukasika kana paine chiitiko chisina kujairika chinowanikwa panguva yekuvhiya.
●Mudziyo uyu haufanirwe kuburitswa munzvimbo ine hunyoro.
●Chengetedza kapu yemafuta nemagetsi zvakachena.Zadza iyokapu nemafuta matsva etransformer kudzivirira kana isingaite.Tarisa kureba kwe electrode uye tarisa kusimba pakati pe electrode tip uye electrode bar screw thread kapu isati yashandiswa zvakare.
1. Oiri Cup Yekuchenesa Nzira
1.1 Pukuta nzvimbo dze electrode uye mbariro zvakare uye zvakare nemucheka wesilika wakachena.
1.2 Gadzirisa kureba kwe electrode nejeji yakajairwa
1.3 Shandisa petroleum ether (mamwe organic solvents anorambidzwa) kuchenesa katatu.Nguva imwe neimwe inofanira kutevera nzira dzinotevera:
① Dururira petroleum ether mukapu yemafuta kusvika kapu yazara 1/4~1/3.
② Fukidza rimu remukombe nechidimbu chegirazi chakacheneswa nepetroleum ether.Zunza mukombe zvakaenzanirana kweminiti imwe neimwe simba.
③ Dururira petroleum ether woomesa kapu neinofuridza kwemaminetsi maviri ~ matatu.
1.4 Shandisa sampu yemafuta kuti iedzwe kuchenesa kapu kwe1 ~ 3 nguva.
① Dururira petroleum ether mukapu yemafuta kusvika kapu yazara 1/4~1/3.
② Fukidza rimu remukombe nechidimbu chegirazi chakacheneswa nepetroleum ether.Zunza mukombe zvakaenzanirana kweminiti imwe neimwe simba.
③ Dururira sampu yeoiri kuruboshwe uye bvunzo inotanga.
2. Kukurudzira tsvimbo yekuchenesa nzira
2.1 Pukuta tsvimbo inobvongodza zvakare uye zvakare nemucheka wesilika wakachena kusvikira zvimedu zvakanaka zvisingawanikwi panzvimbo yavo.Zvinorambidzwa kubata nzvimbo nemaoko.
2.2 Shandisa maforce kusungirira tsvimbo;zviise mupetroleum ether wogeza.
2.3 Shandisa maforce kusungirira tsvimbo nekuiomesa ne blower.
2.4 Shandisa maforce kusungirira tsvimbo;uzviise mumuenzaniso wemafuta wosuka.
3. Oil Cup Storage
Method 1 Zadza kapu nemafuta akanaka ekudzivirira mushure mekuedzwa kwapera uye uise yakagadzikana.
Method 2 Geza uye woomesa mukombe pasi pemaitiro ari pamusoro uye woiisa muvacuum dryer.
Cherechedza: Mukombe wemafuta uye tsvimbo inomutsa ichacheneswa pasi pemaitiro ari pamusoro apa mushure mekuedzwa kwekutanga uye kuedza nemafuta asina kunaka.
4. Common Fault Clearances
4.1 Simba mwenje wakadzimwa, hapana kuratidzwa pachiratidziri
① Tarisa kuti plug yemagetsi yakaiswa zvakasimba kana kwete.
② Tarisa fuse mukati memagetsi emagetsi iri mumamiriro akanaka kana kwete.
③ Tarisa magetsi esoketi.
4.2 Kapu yemafuta isina kuparara chiitiko
① Tarisa kuiswa kwezvibatanidza pabhodhi redunhu.
② Tarisa kubata kweiyo yakakwira-voltage switch pane kavha kavha.
③ Tarisa mabatiro eakakwira-voltage contacts.
④ Tarisa kuputsika kweiyo high-voltage line.
4.3 Musiyano wekuratidzira hauna kukwana
Gadzirisa potentiometer pabhodhi redunhu.
4.4 Printer kutadza
① Tarisa simba replug reprinta.
② Tarisa kuvharwa kweprinta data line.